標題: The Rectenna Design on Contact Lens for Wireless Powering of the Active Intraocular Pressure Monitoring System
作者: Cheng, H. W.
Jeng, B. M.
Chen, C. Y.
Huang, H. Y.
Chiou, J. C.
Luo, C. H.
交大名義發表
National Chiao Tung University
公開日期: 2013
摘要: This paper proposed a wireless power harvesting system with micro-electro-mechanical-systems (MEMS) fabrication for noninvasive intraocular pressure (IOP) measurement on soft contact lens substructure. The power harvesting IC consists of a loop antenna, an impedance matching network and a rectifier. The proposed IC has been designed and fabricated by CMOS 0.18um process that operates at the ISM band of 5.8 GHz. The antenna and the power harvesting IC would be bonded together by using flip chip bonding technologies without extra wire interference. The circuit utilized an impedance transformation circuit to boost the input RF signal that improves the circuit performance. The proposed design achieves an RF-to-DC conversion efficiency of 35% at 5.8 GHz.
URI: http://hdl.handle.net/11536/135362
ISBN: 978-1-4577-0216-7
ISSN: 1557-170X
期刊: 2013 35TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC)
起始頁: 3451
結束頁: 3454
Appears in Collections:Conferences Paper