標題: | Novel non-bouncing PM2.5 impactor modified from well impactor ninety-six |
作者: | Le, Thi-Cuc Tsai, Chuen-Jinn 環境工程研究所 Institute of Environmental Engineering |
公開日期: | 1-一月-2017 |
摘要: | The impaction well containing an oil-soaked glass fiber (GF) filter of the well impactor ninety-six (WINS) needs to be replaced every five sampling days to avoid particle overloading which affects the cutoff diameter (d(pa50)) and the sampling accuracy. This study developed a novel modified WINS (M-WINS) which uses water to wash the wetted GF filter substrate clean continuously to eliminate the particle overloading effect and extend the service life of the impactor without the need of impaction well cleaning and replacement of the oil-soaked GF filter. The laboratory test showed that d(pa50) of the M-WINS only varied slightly from 2.44 to 2.49 mu m while d(pa50) of the WINS decreased from 2.44 to 2.05 mu m when the loaded particle mass increased from 0-6 mg. The field test showed that d(pa50) and the sampling bias of the novel M-WINS met the USEPA requirements (d(pa50) = 2.5 +/- 0.2 mu m and sampling bias +/- 5%) while the WINS without regular replacement of the well showed the decrease in d(pa50) down to 2.12 mu m and the increase in the negative sampling bias to -20% after 17 continuous sampling days. Therefore, the current M-WINS can be used as the PM2.5 inlet for unattended, continuous sampling for a long time with a very good sampling accuracy. |
URI: | http://dx.doi.org/10.1080/02786826.2017.1341621 http://hdl.handle.net/11536/144157 |
ISSN: | 0278-6826 |
DOI: | 10.1080/02786826.2017.1341621 |
期刊: | AEROSOL SCIENCE AND TECHNOLOGY |
Volume: | 51 |
起始頁: | 1287 |
結束頁: | 1295 |
顯示於類別: | 期刊論文 |