| 標題: | A time-delay calibrated method for cornea hysteresis and intraocular pressure measurement |
| 作者: | Wang, Kuo-Jen Tsai, Che-Liang Wang, Wai Hsu, Long Hsu, Ken-Yuh 電子物理學系 光電工程學系 Department of Electrophysics Department of Photonics |
| 關鍵字: | Tonometer;Intraocular pressure;Cornea hysteresis |
| 公開日期: | 1-Apr-2016 |
| 摘要: | The presence of cornea hysteresis (CH) in characterizing the intraocular pressure (IOP) of a human eye deteriorates the accuracy of IOP. To suppress CH, the pressure gauge of a tonometer must be located as close as possible to the cornea. However, this arrangement is unpractical because appropriate working distance to the cornea is required. In this paper, a time-delay calibrated (TDC) method is proposed to counteract the undesired effect of CH in characterizing the IOP. Employing this TDC method, the CH approaches to zero for most eyes measured. |
| URI: | http://dx.doi.org/10.1007/s10043-015-0168-4 http://hdl.handle.net/11536/145495 |
| ISSN: | 1340-6000 |
| DOI: | 10.1007/s10043-015-0168-4 |
| 期刊: | OPTICAL REVIEW |