標題: A time-delay calibrated method for cornea hysteresis and intraocular pressure measurement
作者: Wang, Kuo-Jen
Tsai, Che-Liang
Wang, Wai
Hsu, Long
Hsu, Ken-Yuh
電子物理學系
光電工程學系
Department of Electrophysics
Department of Photonics
關鍵字: Tonometer;Intraocular pressure;Cornea hysteresis
公開日期: 1-Apr-2016
摘要: The presence of cornea hysteresis (CH) in characterizing the intraocular pressure (IOP) of a human eye deteriorates the accuracy of IOP. To suppress CH, the pressure gauge of a tonometer must be located as close as possible to the cornea. However, this arrangement is unpractical because appropriate working distance to the cornea is required. In this paper, a time-delay calibrated (TDC) method is proposed to counteract the undesired effect of CH in characterizing the IOP. Employing this TDC method, the CH approaches to zero for most eyes measured.
URI: http://dx.doi.org/10.1007/s10043-015-0168-4
http://hdl.handle.net/11536/145495
ISSN: 1340-6000
DOI: 10.1007/s10043-015-0168-4
期刊: OPTICAL REVIEW
Volume: 23
起始頁: 340
結束頁: 345
Appears in Collections:Articles