標題: Mathematical Derivation, Circuits Design and Clinical Experiments of Measuring Blood Flow Volume (BFV) at Arteriovenous Fistula (AVF) of Hemodialysis (HD) Patients Using a Newly-Developed Photoplethysmography (PPG) Sensor
作者: Paul, Paul C. P.
Chiang, Pei-Yu
Tarng, D. C.
Yang, C. Y.
電子工程學系及電子研究所
Department of Electronics Engineering and Institute of Electronics
關鍵字: Photoplethysmography (PPG) Sensor;Arteriovenous Fistula (AVF);Blood Flow Volume (BFV)
公開日期: 1-一月-2018
摘要: Mathematical derivation of calculating blood flow volume (BFV) at arteriovenous fistula (AVF) using a newly-developed photoplethysmography (PPG) sensor is presented in this work. Also, the readout circuit of the PPG sensor intended to increase the signal-noise ratio (S/NR) is designed and presented in this work. The designed PPG sensor equipped with derived mathematical equations shows high correlation (R-2 = 0.7563) and low error (RMSE = 212 ml/min) as compared to the gold standard.
URI: http://hdl.handle.net/11536/146169
ISSN: 1948-3287
期刊: 2018 19TH INTERNATIONAL SYMPOSIUM ON QUALITY ELECTRONIC DESIGN (ISQED)
起始頁: 353
結束頁: 356
顯示於類別:會議論文