標題: | Mathematical Derivation, Circuits Design and Clinical Experiments of Measuring Blood Flow Volume (BFV) at Arteriovenous Fistula (AVF) of Hemodialysis (HD) Patients Using a Newly-Developed Photoplethysmography (PPG) Sensor |
作者: | Paul, Paul C. P. Chiang, Pei-Yu Tarng, D. C. Yang, C. Y. 電子工程學系及電子研究所 Department of Electronics Engineering and Institute of Electronics |
關鍵字: | Photoplethysmography (PPG) Sensor;Arteriovenous Fistula (AVF);Blood Flow Volume (BFV) |
公開日期: | 1-一月-2018 |
摘要: | Mathematical derivation of calculating blood flow volume (BFV) at arteriovenous fistula (AVF) using a newly-developed photoplethysmography (PPG) sensor is presented in this work. Also, the readout circuit of the PPG sensor intended to increase the signal-noise ratio (S/NR) is designed and presented in this work. The designed PPG sensor equipped with derived mathematical equations shows high correlation (R-2 = 0.7563) and low error (RMSE = 212 ml/min) as compared to the gold standard. |
URI: | http://hdl.handle.net/11536/146169 |
ISSN: | 1948-3287 |
期刊: | 2018 19TH INTERNATIONAL SYMPOSIUM ON QUALITY ELECTRONIC DESIGN (ISQED) |
起始頁: | 353 |
結束頁: | 356 |
顯示於類別: | 會議論文 |