完整後設資料紀錄
DC 欄位語言
dc.contributor.authorFujisawa, Takumien_US
dc.contributor.authorNoda, Toshihikoen_US
dc.contributor.authorHayashi, Megumien_US
dc.contributor.authorKobe, Ryoen_US
dc.contributor.authorTashiro, Hiroyukien_US
dc.contributor.authorTakehara, Hiroakien_US
dc.contributor.authorSasagawa, Kiyotakaen_US
dc.contributor.authorTokuda, Takashien_US
dc.contributor.authorWu, Chung-Yuen_US
dc.contributor.authorOhta, Junen_US
dc.date.accessioned2019-04-02T06:00:00Z-
dc.date.available2019-04-02T06:00:00Z-
dc.date.issued2016-01-01en_US
dc.identifier.issn0914-4935en_US
dc.identifier.urihttp://hdl.handle.net/11536/147845-
dc.description.abstractHigh-performance electrodes intended for retinal prosthesis were fabricated and evaluated. Iridium oxide (IrOx) was introduced as a high-performance material. A three-dimensional (3D) structure was also introduced to enlarge the electrode's surface area. We tried to improve the electrode performance by combining these approaches, even if the electrode was miniaturized. The effectiveness of IrOx was demonstrated through electrochemical evaluation by comparing it with Pt. IrOx showed 1.6-6 times higher performance for the injection of stimulus pulse current than Pt. The performance of the 3D electrode compared with a planar electrode was also evaluated. Accordingly, the 3D electrode showed 2-4 times higher performance than the planar electrode by surface area enlargement. An ex vivo validation of the stimulus performance was conducted to demonstrate its practical use. A fabricated electrode was implanted in an extracted pig eyeball and the electrochemical performance was evaluated. The fabricated electrode showed sufficient performance of the retinal stimulation, with a high margin of safety. The proposed approach was successfully demonstrated as a stimulus electrode candidate for use in next-generation retinal prosthesis.en_US
dc.language.isoen_USen_US
dc.subjectretinal prosthesisen_US
dc.subjectstimulus electrodeen_US
dc.subjectiridium oxideen_US
dc.subjectthree-dimensional electrodeen_US
dc.subjectcharge injection capacityen_US
dc.titlePerformance Improvement of a Micro-stimulus Electrode for Retinal Prosthesis by Introducing a High-Performance Material and a Three-Dimensional Structureen_US
dc.typeArticleen_US
dc.identifier.journalSENSORS AND MATERIALSen_US
dc.citation.volume28en_US
dc.citation.spage1303en_US
dc.citation.epage1315en_US
dc.contributor.department交大名義發表zh_TW
dc.contributor.departmentNational Chiao Tung Universityen_US
dc.identifier.wosnumberWOS:000392258400007en_US
dc.citation.woscount4en_US