標題: Design and Validation of a New PPG Module to Acquire High-Quality Physiological Signals for High-Accuracy Biomedical Sensing
作者: Kao, Yung-Hua
Chao, Paul C-P
Wey, Chin-Long
電子工程學系及電子研究所
Department of Electronics Engineering and Institute of Electronics
關鍵字: Photoplethysmography (PPG) Sensor;LED/PD PPG module;blood pressure (BP) measurement;reflective pules transient time (R-PTT)
公開日期: 1-一月-2019
摘要: A new photoplethysmography (PPG) module is designed, optimized, and fabricated in this study to obtain quality PPG signal for high-accuracy bio-sensing. The module contains four light emitting diodes (LEDs) in wavelengths of 530, 660, 850, and 940 nm and a photo-diode (PD). The distances between LEDs and PD are optimized for quality PPG signals via maximizing the ratios of pulsatile (ac) to non-pulsatile (dc) components in PPG waveforms. The optimization is carried out based on an establishment of a complete optical model that simulates well optics under the user's skin based on Beer-Lambert law. In results, the optimal LED/PD distances that lead to maximal ac/dc ratios of PPGs for different wavelengths are derived. With the optimal LEDs/PD distances in hand, a new PPG module is designed and fabricated subsequently, and further installed in a hand-held blood pressure (BP) sensing device for performance validation. Based on experimental results, for the PPG module, an ac/dc ratio of 8.02% is achieved by the proposed PPG module as opposed to 6.74% and 6.90% by two other commercial modules. As for sensing BP based on reflective pules transient time, the error is controlled well with the mean difference (MD) and standard deviation (SD) as -1.16 +/- 5.38 (MD +/- SD) and -0.53 +/- 1.92 mmHg, respectively, for systolic blood pressure and diastolic blood pressure. The resulted accuracy successfully categories the developed PPG sensor among the bests by Association for the Advancement of Medical Instrumentation and British Hypertension Society.
URI: http://dx.doi.org/10.1109/JSTQE.2018.2871604
http://hdl.handle.net/11536/148351
ISSN: 1077-260X
DOI: 10.1109/JSTQE.2018.2871604
期刊: IEEE JOURNAL OF SELECTED TOPICS IN QUANTUM ELECTRONICS
Volume: 25
顯示於類別:期刊論文