標題: | A fast digital chip implementing a real-time noise-resistant algorithm for estimating blood pressure using a non-invasive, cuffless PPG sensor |
作者: | Wang, Jeremy H. -S. Yeh, Ming-Hua Chao, Paul C. -P. Tu, Tse-Yi Kao, Yung-Hua Pandey, Rajeev 交大名義發表 電子工程學系及電子研究所 National Chiao Tung University Department of Electronics Engineering and Institute of Electronics |
公開日期: | 1-Jan-1970 |
摘要: | A fast digital chip is designed and fabricated successfully to implement the algorithm for estimating blood pressure (BP) in real time by a non-invasive cuffless photoplethysmogram (PPG) Sensor. The fast response accomplished by this digital computation chip is made possible by one-time calibration only and the Taiwan Semiconductor Manufacturing (TSMC) T18 process to realize the chip in an area of 3.97 mm(2). The experimental result shows that the overall power consumption of the chip is 15.62 mW. Finally, the blood pressure measurement platform has been developed with a user interface (UI) for long-term continuous cuffless BP measurement. The non-invasive blood pressure sensor is applied to the wrist arteries of 44 subjects for sensing the PPG pulsation of the blood vessel, and then estimating BPs (blood pressures). Measurement results shows that the maximum error in the BP measurement is +/- 6 mmHg. Which is less than 8 mmHg, defined by the Advancement of Medical Instrumentation (AAMI) and British Hypertension Society (BHS) standard. |
URI: | http://dx.doi.org/10.1007/s00542-020-04946-y http://hdl.handle.net/11536/154822 |
ISSN: | 0946-7076 |
DOI: | 10.1007/s00542-020-04946-y |
期刊: | MICROSYSTEM TECHNOLOGIES-MICRO-AND NANOSYSTEMS-INFORMATION STORAGE AND PROCESSING SYSTEMS |
起始頁: | 0 |
結束頁: | 0 |
Appears in Collections: | Articles |