標題: Implantable Stimulator for Epileptic Seizure Suppression With Loading Impedance Adaptability
作者: Lin, Chun-Yu
Chen, Wei-Ling
Ker, Ming-Dou
電子工程學系及電子研究所
生醫電子轉譯研究中心
Department of Electronics Engineering and Institute of Electronics
Biomedical Electronics Translational Research Center
關鍵字: Adaptability;epilepsy;epileptic seizure suppression;implantable device;stimulator
公開日期: 1-四月-2013
摘要: The implantable stimulator for epileptic seizure suppression with loading impedance adaptability was proposed in this work. The stimulator consisted of the high voltage generator, output driver, adaptor, and switches, can constantly provide the required 40-mu A stimulus currents, as the loading impedance varied within 10 k Omega-300 k Omega. The performances of this design have been successfully verified in silicon chip fabricated by a 0.35-mu m 3.3-V/24-V CMOS process. The power consumption of this work was only 1.1 mW-1.4 mW. The animal test results with the fabricated chip of proposed design have successfully verified in the Long-Evans rats with epileptic seizures.
URI: http://dx.doi.org/10.1109/TBCAS.2012.2200481
http://hdl.handle.net/11536/21938
ISSN: 1932-4545
DOI: 10.1109/TBCAS.2012.2200481
期刊: IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS
Volume: 7
Issue: 2
起始頁: 196
結束頁: 203