標題: | Trace determination of nitrated polycyclic aromatic hydrocarbons using liquid chromatography with on-line electrochemical reduction and fluorescence detection |
作者: | Kuo, CT Chen, HW Lin, ST 環境工程研究所 Institute of Environmental Engineering |
關鍵字: | nitrated polycyclic aromatic hydrocarbons;ultrasonic extraction;high performance liquid chromatography;electrochemical reduction;fluorescence detection |
公開日期: | 15-Apr-2003 |
摘要: | An efficient clean-up procedure coupled with a high performance liquid chromatography (HPLC) with on-line electrochemical (EC) reduction and fluorescence detection (FLD) was developed to quantify nitrated polycyclic aromatic hydrocarbons (NPAHs) in the airborne particulate. In this process, NPAHs were extracted ultrasonically followed by analysis by using a reversed phase column with an aqueous eluent containing 70% aqueous acetonitrile and sodium monocholoroacetate as a buffer solution. The extraction efficiencies were above 83% for 1-nitropyrene and 1,3-dinitropyrene (1,3-DNP) 1,6-DNP, and 1,8-DNP, and calibration graphs were linear with very good correlation coefficients (r > 0.999) and the detection limits were in the range of 1.0-2.2 pg for dinitropyrenes and nitropyrene. The proposed method provides a relatively simple and convenient procedure for determining the NPAHs samples in airborne particulate. (C) 2003 Elsevier Science B.V. All rights reserved. |
URI: | http://dx.doi.org/10.1016/S0003-2670(03)00204-6 http://hdl.handle.net/11536/27958 |
ISSN: | 0003-2670 |
DOI: | 10.1016/S0003-2670(03)00204-6 |
期刊: | ANALYTICA CHIMICA ACTA |
Volume: | 482 |
Issue: | 2 |
起始頁: | 219 |
結束頁: | 228 |
Appears in Collections: | Articles |
Files in This Item:
If it is a zip file, please download the file and unzip it, then open index.html in a browser to view the full text content.