標題: | Optimization of the headspace solid-phase microextraction for determination of glycol ethers by orthogonal array designs |
作者: | Huang, CT Su, YY Hsieh, YZ 應用化學系 Department of Applied Chemistry |
關鍵字: | orthogonal array designs;solid-phase microextraction;headspace analysis;glycol ethers |
公開日期: | 15-十一月-2002 |
摘要: | A headspace solid-phase microextraction (HS-SPME), in conjunction with gas chromatography-flame ionization detection for use in the determination of six frequently used glycol ethers at the mug/l level is described. A 75 mum Carboxen-polydimethylsiloxane fiber was used to extract the analytes from an aqueous solution. Experimental HS-SPME parameters such as extraction temperature, extraction time, salt concentration and sample volume, were investigated and optimized by orthogonal array experimental designs. The relative standard deviations for the reproducibility of the optimized HS-SPME method varied from 1.48 to 7.59%. The correlation coefficients of the calibration curves exceeded 0.998 in the mug/l range of concentration with at least two orders of magnitude. The method detection limits for glycol ethers in deionized water were in the range of 0.26 to 3.42 mug/l. The optimized method was also applied to the analysis of glycol ethers in urine and blood samples with the method detection limits ranged from 1.74 to 23.2 mug/l. (C) 2002 Elsevier Science B.V. All rights reserved. |
URI: | http://dx.doi.org/10.1016/S0021-9673(02)01278-5 http://hdl.handle.net/11536/28379 |
ISSN: | 0021-9673 |
DOI: | 10.1016/S0021-9673(02)01278-5 |
期刊: | JOURNAL OF CHROMATOGRAPHY A |
Volume: | 977 |
Issue: | 1 |
起始頁: | 9 |
結束頁: | 16 |
顯示於類別: | 期刊論文 |