标题: 应用具互补式金氧半感测晶片之生医感测系统于三酸甘油脂检测
The Detection of Triglyceride using a Biosensing System with a CMOS Sensor Chip
作者: 蔡翼钟
Tsai, I-Chung
黄远东
Huang, Yang-Tung
电子研究所
关键字: 三酸甘油脂;互补式金氧半;感测晶片;生医感测;醌亚胺;Triglyceride;CMOS;Sensor Chip;Biosensing;Quinoneimine Dye
公开日期: 2012
摘要: 本研究探讨应用具互补式金氧半感测晶片之生医感测系统于三酸甘油脂检测。由于代谢症候群已成为常见的文明病之一,而血液中的三酸甘油脂含量又是临床检测代谢症候群的重要指标,故三酸甘油脂之生医检测系统已成为近代医疗科技发展的趋势之一。三酸甘油脂透过一系列之酵素反应后可生成紫色之醌亚胺(quinoneimine dye),对于绿光之吸收度和三酸甘油脂浓度呈正相关,适合以吸光光度法检测。本生医感测系统即透过光电二极体将光学讯号转成电流讯号,并以转阻放大器将之转换成电压讯号,进而达到侦测发光二极体放射之绿光因醌亚胺吸收而产生的光强度差异。此一系统核心晶片结合前述之光电二极体和转阻放大器,另有一外接回馈电阻,可调整电路放大倍率。该晶片相容于互补式金氧半导体制程,具有体积小、成本低及可大量制造等特性,使用方便且易于普及。实验结果显示,此生医感测系统在0.1 – 2.5 mg/mL范围内能定量血清样本中三酸甘油脂之浓度,同时证实ABTS/H2O2/HRP反应可应用于三酸甘油脂于全血样本中的检测,能定量三酸甘油脂稀释后浓度小于7.3 μg/mL的全血样本,符合临床检验需求。
In this research, the detection of triglyceride using a biosensing system with a CMOS sensor chip is investigated. In the past few decades, metabolic syndrome has become more and more common in the developed countries. Since triglyceride is one of the key traits to determine metabolic syndrome in the clinical diagnosis, to develop a triglyceride-detecting biosensing system has become a trend in the health care industry. While the absorbance of the green light passing through the purple quinoneimine dye, which is produced after a series of biochemical reactions involving enzymatic hydrolysis of the triglyceride, is proportional to the concentration of triglyceride, the absorption photometry is adopted in this study. A photodiode in the biosensing system converts the optical signal to the current signal first, and the a transimpedance amplifier (TIA) converts it to the voltage signal, such that the absorbance of the green light emitted from an LED can be detected. A CMOS sensor biochip integrated the photodiode and the TIA with an external feedback resistor to adjust the sensitivity. The CMOS chip makes the biosensing system portable, inexpensive, and easy-to-use. In summary, the experimental results have shown that the biosensing system can be applied to detect the concentration of triglyceride in serum samples within the range of 0.1 – 2.5 mg/mL. The ABTS/H2O2/HRP reaction was also adopted to detect the triglyceride concentration below 7.3 μg/mL in a diluted whole blood sample. Both of methods are suitable for the clinical diagnosis.
URI: http://140.113.39.130/cdrfb3/record/nctu/#GT079911533
http://hdl.handle.net/11536/49081
显示于类别:Thesis