A 0.09 mu W Low Power Front-End Biopotential Amplifier for Biosignal Recording

Loading...
Thumbnail Image

Journal Title

Journal ISSN

Volume Title

Publisher

DOI

10.1109/TBCAS.2012.2188029

Abstract

This work presents a biopotential front-end amplifier in which the MOS transistors are biased in subthreshold region with a supply voltage and current of 0.4-0.8 V and 0.23-1.86 mu A, respectively, to reduce the system power. Flicker noise is then removed using a chopping technique, and differential interference produced by electrode impedance imbalance is suppressed using a Gm-C filter. Additionally, the circuit is fabricated using TSMC 0.18 mu m CMOS technology with a core area of 0.77 X 0.36 mm(2). With a minimum supply voltage of 0.4 V, the measured SNR and power consumption of the proposed IC chip are 54.1 dB and 0.09 mu W, respectively.