Label-free imaging of melanoma with nonlinear photothermal microscopy

Loading...
Thumbnail Image

Journal Title

Journal ISSN

Volume Title

Publisher

DOI

10.1364/OL.40.001141

Abstract

Nonlinear photothermal microscopy, in which the intensity of the pump heating beam is modulated at f and the photothermal signal is extracted from the probe beam with a lock-in amplifier referred to 2f, is applied to the imaging of mouse melanoma without any staining. The pump and probe pulses, with central wavelengths of 488 and 632 nm, and a pulse duration of similar to 100 ps, are filtered from a compact commercial supercontinuum fiber laser source. An auto-balanced detector is applied to accumulate the signal and remove the laser noise of the probe. The spatial resolution of the nonlinear photothermal imaging is enhanced by similar to 18% in both theoretical calculations and experiments, compared with a linear photothermal mechanism, and the resolution enhancement is theoretically similar to 42% compared with conventional optical microscopy. This imaging technique shows possibilities for the clinical evaluation of melanoma with a high contrast and spatial resolution. (C) 2015 Optical Society of America