Abstract

Using first-order treatment on azimuth errors of the polarizer (P) and analyzer (A) with respect to the plane of incidence (POI), we construct two intensity ratios around a certain configuration: such as P = +/- pi/4 and A = 0 as compared to A = pi/2. These expressions can be used not only to align the azimuths of the polarizer and analyzer to the POI, hut also to obtain the ellipsometric parameter phi without locating the minimum intensity, BK7 glass and SiO2/Si thin film are ultilized to evaluate this technique.