標題: A 16-Channel CMOS Chopper-Stabilized Analog Front-End ECoG Acquisition Circuit for a Closed-Loop Epileptic Seizure Control System
作者: Wu, Chung-Yu
Cheng, Cheng-Hsiang
Chen, Zhi-Xin
電子工程學系及電子研究所
生醫電子轉譯研究中心
Department of Electronics Engineering and Institute of Electronics
Biomedical Electronics Translational Research Center
關鍵字: Closed-loop seizure control;chopper-stabilized;input protection circuit;low noise;low power;neural-signal amplifier;offset reduction loop;ripple reduction loop
公開日期: 1-六月-2018
摘要: In this paper, a 16-channel analog front-end (AFE) electrocorticography signal acquisition circuit for a closed-loop seizure control system is presented. It is composed of 16 input protection circuits, 16 auto-reset chopper-stabilized capacitive-coupled instrumentation amplifiers (AR-CSCCIA) with bandpass filters, 16 programmable transconductance gain amplifiers, a multiplexer, a transimpedance amplifier, and a 128-kS/s 10-bit delta-modulated successive-approximation-register analog-todigital converter (SAR ADC). In closed-loop seizure control system applications, the stimulator shares the same electrode with the AFE amplifier for effective suppression of epileptic seizures. To prevent from overstress in MOS devices caused by high stimulation voltage, an input protection circuit with a high-voltage-tolerant switch is proposed for the AFE amplifier. Moreover, low input-referred noise is achieved by using the chopper modulation technique in the AR-CSCCIA. To reduce the undesired effects of chopper modulation, an improved offset reduction loop is proposed to reduce the output offset generated by input chopper mismatches. The digital ripple reduction loop is also used to reduce the chopper ripple. The fabricated AFE amplifier has 49.1-/59.4-/67.9-dB programmable gain and 2.02-mu Vrms input referred noise in a bandwidth of 0.59-117 Hz. The measured power consumption of the AFE amplifier is 3.26 mu W per channel, and the noise efficiency factor is 3.36. The in vivo animal test has been successfully performed to verify the functions. It is shown that the proposed AFE acquisition circuit is suitable for implantable closed-loop seizure control systems.
URI: http://dx.doi.org/10.1109/TBCAS.2018.2808415
http://hdl.handle.net/11536/145097
ISSN: 1932-4545
DOI: 10.1109/TBCAS.2018.2808415
期刊: IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS
Volume: 12
起始頁: 543
結束頁: 553
顯示於類別:期刊論文