標題: Balance Control and Energetics of Powered Exoskeleton-Assisted Sit-to-Stand Movement in Individuals With Paraplegic Spinal Cord Injury
作者: Mao, Hui-Fen
Huang, Hsing-Po
Lu, Tung-Wu
Wang, Ting-Ming
Wu, Cheng-Hua
Hu, Jwu-Sheng
電控工程研究所
Institute of Electrical and Control Engineering
關鍵字: Exoskeleton device;Postural balance;Rehabilitation;Spinal cord injury
公開日期: 1-十月-2018
摘要: Objective: To quantify the effects of initial hip angle and angular hip velocity settings of a lower-limb wearable robotic exoskeleton (WRE) on the balance control and mechanical energy requirements in patients with paraplegic spinal cord injuries (SCIs) during WRE-assisted sit-to-stand (STS). Design: Observational, cross-sectional study. Setting: A university hospital gait laboratory with an 8-camera motion analysis system, 3 forceplates, a pair of instrumented crutches, and a WRE. Participants: Patients (N = 12) with paraplegic SCI. Interventions: Not applicable. Main Outcome Measures: The inclination angle (IA) of the body's center of mass (COM) relative to the center of pressure (COP), and the rate of change of IA (RCIA) for balance control, and the mechanical energy and forward COM momentum before and after seat-off for energetics during WRE-assisted STS were compared between conditions with 2 initial hip angles (105 degrees and 115 degrees) and 3 initial hip angular velocities (800, 1000, 1200 rpm). Results: No interactions between the main factors (ie, initial hip angle vs angular velocity) were found for any of the calculated variables. Greater initial hip angle helped the patients with SCI move the body forward with increased COM momentum but reduced RCIA (P<.05). With increasing initial angular hip velocity, the IA and RCIA after seat-off (P<.05) increased linearly while total mechanical energy reduced linearly (P<.05). Conclusions: The current results suggest that a greater initial hip angle with smaller initial angular velocity may provide a favorable compromise between momentum transfer and balance of the body for people with SCI during WRE-assisted STS. The current data will be helpful for improving the design and clinical use of the WRE. (C) 2018 by the American Congress of Rehabilitation Medicine
URI: http://dx.doi.org/10.1016/j.apmr.2018.04.004
http://hdl.handle.net/11536/148283
ISSN: 0003-9993
DOI: 10.1016/j.apmr.2018.04.004
期刊: ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION
Volume: 99
起始頁: 1982
結束頁: 1990
顯示於類別:期刊論文