摘要: | Novel anode layer plasma within minimum chamber space was developed for non-contact alignment process. The plasma-treated polyimide (PI) surface showed no particle contamination and no micro-scratches. Surface morphology was investigated by using scanning electron microscope (SEM), an atomic force microscope (AFM), and X-ray photoemission spectroscopy. The different oxygen-to-carbon ratio ([O]/[C] ratio) for XPS spectra indicated a composition change after plasma treatment. Surface pretilt angles were varied from 0 to 2.1 degrees under different plasma exposure times. Finally, a prototype 20.8-in. QXGA IPS-mode gray-scale medical liquid-crystal display was successfully demonstrated with high contrast ratio, excellent uniformity, and wide viewing angle using this new plasma-beam-alignment technique. |
URI: | http://dx.doi.org/10.1889/1.2835038 http://hdl.handle.net/11536/175 |
ISSN: | 1071-0922 |
DOI: | 10.1889/1.2835038 |
期刊: | JOURNAL OF THE SOCIETY FOR INFORMATION DISPLAY |
Volume: | 16 |
Issue: | 1 |
起始頁: | 71 |
結束頁: | 75 |