標題: | The Ge Enhance the Sensitivity for Bio-Sensor |
作者: | Chang, Kow-Ming Kuo, Jiun-Ming Chao, Wen-Chan Liang, Chia-Jung Wu, Heng-Hsin Tzeng, Wen-Hsien Wu, Tzu-liu 電子工程學系及電子研究所 Department of Electronics Engineering and Institute of Electronics |
公開日期: | 2008 |
摘要: | Nanowire is mainly used in biological sensor because it has the high surface-to-volume ratio. In this study, we have successfully fabricated the N-SiGe nanowire with different Ge concentration by side-wall spacer technique respectively. The 3-amino-propyltrime-thoxy-silane (APTS) was used to modify the surface, which can connect the bio-linker. The conductance of SiGe nanowire Increases owing to APTS with positive charge. The his (sulfosuccinimidyl) suberate sodium (BS3) as the bio-linker connects to APTS, and the conductance decreases because of negative charge. Finally, the protein immunoglobulin G (IGG) is linked to BS3, and the conductance reduces for negative charge. Comparing the conductance change; the SiGe nanowire with higher Ge concentration improved the sensitivity. But the over-higher Ge concentration (40%) did not increase the sensitivity; the reason maybe the higher defect appears at the surface as higher Ge concentration. |
URI: | http://hdl.handle.net/11536/1797 |
ISBN: | 978-1-4244-1572-4 |
期刊: | 2008 2ND IEEE INTERNATIONAL NANOELECTRONICS CONFERENCE, VOLS 1-3 |
起始頁: | 811 |
結束頁: | 814 |
顯示於類別: | 會議論文 |