Title: The Novel research of Intraocular Pressure Tonometer by Using Inductance sensor
Authors: Tang, Yu-Shun
Jeng, Wei-De
Huang, Ting-Wei
Ou-Yang, Mang
Chiou, Jin-Chern
Duann, Jeng-Ren
Luo, Ching-Hsing
Huang, Hong-Yi
Tsai, Yi-Wu
電控工程研究所
Institute of Electrical and Control Engineering
Issue Date: 2012
Abstract: This article mainly discusses the theoretical derivation and the estimate of the intraocular pressure (IOP) sensor by using inductance. The inductance changes associate with the variation of anterior corneal curvature radius due to the differences of IOP. The value of inductance is transformed by the LC oscillator, and the signal is transmitted by antenna, and finally read by the exterior readout circuit to achieve IOP measuring. The result show that the estimative values of inductance is close to the measured values, and the percentage of inductance variation is about 0.08% per 1 mmHg. Because of the low percentage, it makes much harder to realize readout circuit. In the future, the alteration of structure and material are considered to increase the percentage of varying inductance.
URI: http://hdl.handle.net/11536/21608
ISBN: 978-1-4577-1766-6
ISSN: 1930-0395
Journal: 2012 IEEE SENSORS PROCEEDINGS
Begin Page: 465
End Page: 468
Appears in Collections:Conferences Paper