標題: The Novel research of Intraocular Pressure Tonometer by Using Inductance sensor
作者: Tang, Yu-Shun
Jeng, Wei-De
Huang, Ting-Wei
Ou-Yang, Mang
Chiou, Jin-Chern
Duann, Jeng-Ren
Luo, Ching-Hsing
Huang, Hong-Yi
Tsai, Yi-Wu
電控工程研究所
Institute of Electrical and Control Engineering
公開日期: 2012
摘要: This article mainly discusses the theoretical derivation and the estimate of the intraocular pressure (IOP) sensor by using inductance. The inductance changes associate with the variation of anterior corneal curvature radius due to the differences of IOP. The value of inductance is transformed by the LC oscillator, and the signal is transmitted by antenna, and finally read by the exterior readout circuit to achieve IOP measuring. The result show that the estimative values of inductance is close to the measured values, and the percentage of inductance variation is about 0.08% per 1 mmHg. Because of the low percentage, it makes much harder to realize readout circuit. In the future, the alteration of structure and material are considered to increase the percentage of varying inductance.
URI: http://hdl.handle.net/11536/21608
ISBN: 978-1-4577-1766-6
ISSN: 1930-0395
期刊: 2012 IEEE SENSORS PROCEEDINGS
起始頁: 465
結束頁: 468
Appears in Collections:Conferences Paper