标题: Examination of the temporal effect in a flow injection analysis system using multi-channel absorbance detection
作者: Lin, Shu-Hui
Chang, Kai-Chieh
Chen, Yu-Chun
Yu, Tiing
Pai, Su-Cheng
应用化学系
Department of Applied Chemistry
关键字: Poiseuille effect;Temporal effect;Flow injection analysis;Tailing peaks;Asymmetry factor;Multi-channel detection
公开日期: 22-五月-2009
摘要: In addition to the Poiseuille effect, a so-called temporal effect was proposed recently to elucidate the commonly observed tailing peak signals of flow injection analysis (FIA). A multi-channel absorbance detector was used in this study to obtain the FIA peaks on both the spatial and temporal coordinates. The temporal effect was analyzed by comparison of the profiles between the experimental and the corresponding Gaussian peaks. and by comparison of asymmetry factors between the spatial and the temporal peaks. The temporal effect appeared to be the major factor under flow rates ranging from 0.5 to 8 ml min(-1). This was despite the presence of a spatially frontal peak observed in the FIA tubing, which was found to result in a tailing peak on the temporal coordinate due to this discussed cause. in addition, the temporal effect became greater as the flow rate increased. (C) 2009 Published by Elsevier B.V.
URI: http://dx.doi.org/10.1016/j.chroma.2008.10.073
http://hdl.handle.net/11536/7224
ISSN: 0021-9673
DOI: 10.1016/j.chroma.2008.10.073
期刊: JOURNAL OF CHROMATOGRAPHY A
Volume: 1216
Issue: 21
起始页: 4618
结束页: 4624
显示于类别:Articles