標題: | A Phototransistor-Based High-Sensitivity Biosensing System Using 650-nm Light |
作者: | Chang, Yu-Wei Tai, Yu-Ting Huang, Yang-Tung Yang, Yuh-Shyong 生物科技學系 電子工程學系及電子研究所 Department of Biological Science and Technology Department of Electronics Engineering and Institute of Electronics |
關鍵字: | Biomedical transducers;medical diagnosis;photodetectors;phototransistors |
公開日期: | 1-May-2009 |
摘要: | A miniature optical biosensing system based on a PMOS phototransistor and absorption photometry is proposed. The phototransistor was manufactured in a standard 0.35-mu m CMOS process, and it exhibited a responsivity higher than 1000 A/W for 650-nm light. For biochemical applications, the TMB/H(2)O(2)/HRP method was adopted as a useful basis in our system. A sample volume of only 10 mu l was required to be dropped on the slide above the phototransistor. Experimental results demonstrated that a high sensitivity of 2.5 mu A/pM was achieved, and the minimum HRP concentration successfully detected was 2.7 pM. This detection limit is three orders of magnitude better than that of a lately reported silicon biosensor, and is even comparable to that of a commercial spectrophotometer. |
URI: | http://dx.doi.org/10.1109/JSEN.2009.2020675 http://hdl.handle.net/11536/7315 |
ISSN: | 1530-437X |
DOI: | 10.1109/JSEN.2009.2020675 |
期刊: | IEEE SENSORS JOURNAL |
Volume: | 9 |
Issue: | 5-6 |
起始頁: | 673 |
結束頁: | 677 |