標題: | Novel Wearable and Wireless Ring-Type Pulse Oximeter with Multi-Detectors |
作者: | Huang, Cheng-Yang Chan, Ming-Che Chen, Chien-Yue Lin, Bor-Shyh 影像與生醫光電研究所 Institute of Imaging and Biomedical Photonics |
關鍵字: | pulse oximeter;arterial oxygen saturation;finger base;optical human tissue simulation;multi-detector |
公開日期: | 1-九月-2014 |
摘要: | The pulse oximeter is a popular instrument to monitor the arterial oxygen saturation (SPO2). Although a fingertip-type pulse oximeter is the mainstream one on the market at present, it is still inconvenient for long-term monitoring, in particular, with respect to motion. Therefore, the development of a wearable pulse oximeter, such as a finger base-type pulse oximeter, can effectively solve the above issue. However, the tissue structure of the finger base is complex, and there is lack of detailed information on the effect of the light source and detector placement on measuring SPO2. In this study, the practicability of a ring-type pulse oximeter with a multi-detector was investigated by optical human tissue simulation. The optimal design of a ring-type pulse oximeter that can provide the best efficiency of measuring SPO2 was discussed. The efficiency of ring-type pulse oximeters with a single detector and a multi-detector was also discussed. Finally, a wearable and wireless ring-type pulse oximeter was also implemented to validate the simulation results and was compared with the commercial fingertip-type pulse oximeter. |
URI: | http://dx.doi.org/10.3390/s140917586 http://hdl.handle.net/11536/25382 |
ISSN: | 1424-8220 |
DOI: | 10.3390/s140917586 |
期刊: | SENSORS |
Volume: | 14 |
Issue: | 9 |
起始頁: | 17586 |
結束頁: | 17599 |
顯示於類別: | 期刊論文 |