標題: | Heterodyne moire interferometry for measuring corneal surface profile |
作者: | Chang, Wei-Yao Chen, Kun-Huang Chen, Der-Chin Tseng, Jung-Kai Chen, Shyan-Tarng Sun, Han-Ying Chen, Jing-Heng Hsu, Ken Y. 光電工程學系 Department of Photonics |
關鍵字: | Corneal surface profile;Heterodyne interferometry;Projection moire;Talbot effect |
公開日期: | 1-Mar-2014 |
摘要: | This study proposes an accurate method for reconstructing the corneal surface profile. By applying a constant velocity to the projection grating along the grating plane, a series of sampling points of the sinusoidal wave, which behaves in the manner of heterodyne interferometric signals, can be recorded using a CMOS camera. The phase distribution of the moire fringes can then be obtained using the IEEE 1241 least-square sine fitting algorithm and two-dimensional (2D) phase unwrapping. Finally, the corneal surface profile can be reconstructed by substituting the phase distribution into a specially derived equation. To validate the proposed method, the corneal surface of a pig eyeball was measured. The measurement resolution was approximately 3.5 mu m. Because of the introduction of the Talbot effect, the projection moire method, and heterodyne interferometry, this approach provides the advantages of a simple optical setup, ease of operation, high stability, and high resolution. (C) 2013 Elsevier Ltd. All rights reserved. |
URI: | http://dx.doi.org/10.1016/j.optlaseng.2013.07.013 http://hdl.handle.net/11536/23366 |
ISSN: | 0143-8166 |
DOI: | 10.1016/j.optlaseng.2013.07.013 |
期刊: | OPTICS AND LASERS IN ENGINEERING |
Volume: | 54 |
Issue: | |
起始頁: | 232 |
結束頁: | 235 |
Appears in Collections: | Articles |
Files in This Item:
If it is a zip file, please download the file and unzip it, then open index.html in a browser to view the full text content.