標題: | Mixed Bio-Signal Lossless Data Compressor for Portable Brain-Heart Monitoring Systems |
作者: | Chua, Ericson Fang, Wai-Chi 電子工程學系及電子研究所 Department of Electronics Engineering and Institute of Electronics |
關鍵字: | Biomedical signal processing;lossless data compression;electroencephalogram (EEG);electrocardiogram (ECG);diffuse optical tomography (DOT) |
公開日期: | 1-Feb-2011 |
摘要: | This paper presents a highly integrated VLSI implementation of a mixed bio-signal lossless data compressor capable of handling multichannel electroencephalogram (EEG), electrocardiogram (ECG) and diffuse optical tomography (DOT) bio-signal data for reduced storage and communication bandwidth requirements in portable, wireless brain-heart monitoring systems used in hospital or home care settings. The compressor integrated in a multiprocessor brain-heart monitoring IC comprises 15k gates and 12kbits of RAM, occupying a total area of 58k mu m(2) in 65nm CMOS technology. Results demonstrate an average compression ratio (CR) of 2.05, and a simulated power consumption of 170 mu W at an operating condition of 24MHz clock and 1.0V core voltage. Nominal power savings of 43% and 47% at the transmitter can be achieved when employing Bluetooth and Zigbee transceivers, respectively(1). |
URI: | http://dx.doi.org/10.1109/TCE.2011.5735512 http://hdl.handle.net/11536/25782 |
ISSN: | 0098-3063 |
DOI: | 10.1109/TCE.2011.5735512 |
期刊: | IEEE TRANSACTIONS ON CONSUMER ELECTRONICS |
Volume: | 57 |
Issue: | 1 |
起始頁: | 267 |
結束頁: | 273 |